1
/
trong số
12
PayPal, credit cards. Download editable-PDF & invoice in 1 second!
YY 0042-2018 English PDF (YY0042-2018)
YY 0042-2018 English PDF (YY0042-2018)
Giá thông thường
$130.00 USD
Giá thông thường
Giá ưu đãi
$130.00 USD
Đơn giá
/
trên
Phí vận chuyển được tính khi thanh toán.
Không thể tải khả năng nhận hàng tại cửa hàng
Delivery: 3 seconds. Download true-PDF + Invoice.
Get QUOTATION in 1-minute: Click YY 0042-2018
Historical versions: YY 0042-2018
Preview True-PDF (Reload/Scroll if blank)
YY 0042-2018: High-frequency jet ventilator
YY 0042-2018
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.10
C 46
Replacing YY 0042-2007
High-Frequency Jet Ventilator
ISSUED ON: DECEMBER 20, 2018
IMPLEMENTED ON: JUNE 01, 2020
Issued by: National Medical Products Administration
Table of Contents
Foreword ... 3
1 Scope ... 5
2 Normative References ... 5
3 Terms and Definitions ... 6
4 Gas-Source Supply ... 6
5 Power Supply ... 7
6 Mechanical Strength ... 8
7 Waterproof ... 8
8 Fireproof ... 8
9 Cleaning, Disinfection or Sterilization ... 8
10 Material Compatibility ... 9
11 Accuracy of Work Data ... 9
12 Prevention of Dangerous Outputs ... 10
13 Connection ... 11
14 Marking ... 11
15 Information Provided by the Manufacturer ... 12
High-Frequency Jet Ventilator
1 Scope
This Standard specifies the basic safety, performance, marking and information provided by the
manufacturer for high-frequency jet ventilator (hereinafter referred to as ventilator).
This Standard applies to high-frequency jet ventilator with both exhalation and inhalation open.
The device is suitable for use by patients who need respiratory support, respiratory therapy and
emergency resuscitation under the supervision of medical staff.
2 Normative References
The following documents are essential to the application of this document. For the dated
documents, only the versions with the dates indicated are applicable to this document; for the
undated documents, only the latest version (including all the amendments) is applicable to this
document.
GB/T 4999 Anesthetic Equipment and Breathing Machine - Vocabulary
GB 9706.1 Medical Electrical Equipment - Part 1: General Requirements for Safety
GB 18278.1 Sterilization of Health Care Products - Moist Heat - Part 1: Requirements for
the Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18279.1 Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements
for Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18280.1 Sterilization of Health Care Products - Radiation - Part 1: Requirements for
Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18281.1 (all parts) Sterilization of Health Care Products - Biological Indicators
YY/T 0466.1-2016 Medical Devices - Symbols to be Used with Medical Device Labels,
Labelling and Information to be Supplied - Part 1: General Requirements
YY 0600.3-2007 Lung Ventilators for Medical Use - Particular Requirements for Basic
Safety and Essential Performance - Part 3: Emergency and Transport Ventilators
YY 0601 Medical Electrical Equipment - Particular Requirements for Basic Safety and
normally within the input pressure range of 280kPa~600kPa. It shall not cause any hazard when
the pressure of the air inlet is up to 1000kPa. When the gas source pressure is 280kPa, the
average gas flow measured at the high-pressure gas input port of the ventilator shall not exceed
60L/min (over 10s time-weighted average); and the instantaneous flow shall not exceed
200L/min (within 3s).
4.2 Alarm of abnormal gas-source supply
The ventilator shall have an alarm function of abnormal gas-source supply. When the gas-
source supply pressure is too low to maintain the normal operation of the ventilator, the
ventilator shall be able to generate a high-priority alarm or a continuous audible alarm in
accordance with YY 0709; and the audible alarm signal shall at least last 120s.
4.3 One gas missing in the air-oxygen mixing system
In addition to the air-oxygen mixing system mixed with the atmosphere, if a gas is missing in
the air-oxygen mixing system, the ventilator shall be able to automatically switch to make the
remaining gas continue working; and at the same time, it shall generate an alarm signal of at
least low priority in accordance with YY 0709.
5 Power Supply
5.1 Alarm on the abnormality of grid power supply
The ventilator that is powered only by the grid power supply shall have the function of alarming
the abnormality of grid power supply. When the voltage of the grid power is too low to maintain
the normal operation of the ventilator, a high-priority alarm or a continuous audible alarm in
accordance with YY 0709 shall be activated; and the audible alarm signal shall last at least 120s.
5.2 DC power supply
In addition to the grid power supply, if the ventilator is additionally equipped with a DC power
supply (internal power supply or external DC power supply), and can automatically switch to
DC power supply when the grid power supply is abnormal to maintain its normal operation,
then when the grid power supply is abnormal, the alarm on the abnormality of grid power supply
shall not be activated. For the conversion of such power supply, it shall comply with the
provisions of the information prompt signal or low-priority alarm signal specified in YY 0709.
The ventilator that is powered by DC power supply should activate a medium-priority alarm in
accordance with YY 0709 when the battery is low; and should activate a high-priority alarm in
accordance with YY 0709 before all power is lost.
If the ventilator is equipped with a DC power supply, a method for the operator to determine
the status of the DC power supply shall be provided.
5.3 Prevent mis-operation of the power switch
A method to prevent mis-operation of the power switch shall be provided.
5.4 Power cord
If the ventilator does not have a backup power supply, its equipped grid power cord shall be a
non-detachable power cord; or measures shall be provided to prevent the grid power cord from
accidentally falling off the ventilator.
6 Mechanical Strength
The ventilator that can be used for out-of-hospital transport shall be able to withstand the stress
caused by rough operation during working; and the ventilator shall continue to work during and
after the vibration impact test according to 21.6a), 21.6b) and 21.6aa) of YY 0600.3-2007.
7 Waterproof
The waterproof rating of ventilators that can be used for out-of-hospital transport shall be IPX4.
8 Fireproof
In normal or single fault conditions, combustible materials shall not be in both of the following
conditions at the same time:
--- the material temperature reaches the minimum ignition temperature;
--- the presence of oxidizing agents.
9 Cleaning, Disinfection or Sterilization
Accessories and components in the ventilator breathing system that may be contaminated with
rebreathing gases shall be removable for cleaning, disinfection, or sterilization.
If the label indicates that the product is sterile, the product should be sterilized by appropriate
and effective methods in accordance with the relevant descriptions in GB 18278.1, GB 18279.1,
GB 18280.1 and GB 18281.
Packaging systems without passing through the sterilized equipment shall maintain product
cleanliness prior to sterilization, use; and minimize the risk of bacterial contamination.
(0.1kPa + 10% of the setting value).
11.6 Sigh
If the ventilator has a sigh function, the tidal volume during sighing shall be no less than 1.5
times the tidal volume during non-sighing.
11.7 Measurement of airway pressure
A device to measure the airway pressure at the patient's connection port shall be provided; and
the accuracy of the pressure measuring device shall be ± (2% full scale + 8% actual reading).
NOTE: The actual measurement can be carried out at any part o...
Get QUOTATION in 1-minute: Click YY 0042-2018
Historical versions: YY 0042-2018
Preview True-PDF (Reload/Scroll if blank)
YY 0042-2018: High-frequency jet ventilator
YY 0042-2018
PHARMACEUTICAL INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.10
C 46
Replacing YY 0042-2007
High-Frequency Jet Ventilator
ISSUED ON: DECEMBER 20, 2018
IMPLEMENTED ON: JUNE 01, 2020
Issued by: National Medical Products Administration
Table of Contents
Foreword ... 3
1 Scope ... 5
2 Normative References ... 5
3 Terms and Definitions ... 6
4 Gas-Source Supply ... 6
5 Power Supply ... 7
6 Mechanical Strength ... 8
7 Waterproof ... 8
8 Fireproof ... 8
9 Cleaning, Disinfection or Sterilization ... 8
10 Material Compatibility ... 9
11 Accuracy of Work Data ... 9
12 Prevention of Dangerous Outputs ... 10
13 Connection ... 11
14 Marking ... 11
15 Information Provided by the Manufacturer ... 12
High-Frequency Jet Ventilator
1 Scope
This Standard specifies the basic safety, performance, marking and information provided by the
manufacturer for high-frequency jet ventilator (hereinafter referred to as ventilator).
This Standard applies to high-frequency jet ventilator with both exhalation and inhalation open.
The device is suitable for use by patients who need respiratory support, respiratory therapy and
emergency resuscitation under the supervision of medical staff.
2 Normative References
The following documents are essential to the application of this document. For the dated
documents, only the versions with the dates indicated are applicable to this document; for the
undated documents, only the latest version (including all the amendments) is applicable to this
document.
GB/T 4999 Anesthetic Equipment and Breathing Machine - Vocabulary
GB 9706.1 Medical Electrical Equipment - Part 1: General Requirements for Safety
GB 18278.1 Sterilization of Health Care Products - Moist Heat - Part 1: Requirements for
the Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18279.1 Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements
for Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18280.1 Sterilization of Health Care Products - Radiation - Part 1: Requirements for
Development, Validation and Routine Control of a Sterilization Process for Medical
Devices
GB 18281.1 (all parts) Sterilization of Health Care Products - Biological Indicators
YY/T 0466.1-2016 Medical Devices - Symbols to be Used with Medical Device Labels,
Labelling and Information to be Supplied - Part 1: General Requirements
YY 0600.3-2007 Lung Ventilators for Medical Use - Particular Requirements for Basic
Safety and Essential Performance - Part 3: Emergency and Transport Ventilators
YY 0601 Medical Electrical Equipment - Particular Requirements for Basic Safety and
normally within the input pressure range of 280kPa~600kPa. It shall not cause any hazard when
the pressure of the air inlet is up to 1000kPa. When the gas source pressure is 280kPa, the
average gas flow measured at the high-pressure gas input port of the ventilator shall not exceed
60L/min (over 10s time-weighted average); and the instantaneous flow shall not exceed
200L/min (within 3s).
4.2 Alarm of abnormal gas-source supply
The ventilator shall have an alarm function of abnormal gas-source supply. When the gas-
source supply pressure is too low to maintain the normal operation of the ventilator, the
ventilator shall be able to generate a high-priority alarm or a continuous audible alarm in
accordance with YY 0709; and the audible alarm signal shall at least last 120s.
4.3 One gas missing in the air-oxygen mixing system
In addition to the air-oxygen mixing system mixed with the atmosphere, if a gas is missing in
the air-oxygen mixing system, the ventilator shall be able to automatically switch to make the
remaining gas continue working; and at the same time, it shall generate an alarm signal of at
least low priority in accordance with YY 0709.
5 Power Supply
5.1 Alarm on the abnormality of grid power supply
The ventilator that is powered only by the grid power supply shall have the function of alarming
the abnormality of grid power supply. When the voltage of the grid power is too low to maintain
the normal operation of the ventilator, a high-priority alarm or a continuous audible alarm in
accordance with YY 0709 shall be activated; and the audible alarm signal shall last at least 120s.
5.2 DC power supply
In addition to the grid power supply, if the ventilator is additionally equipped with a DC power
supply (internal power supply or external DC power supply), and can automatically switch to
DC power supply when the grid power supply is abnormal to maintain its normal operation,
then when the grid power supply is abnormal, the alarm on the abnormality of grid power supply
shall not be activated. For the conversion of such power supply, it shall comply with the
provisions of the information prompt signal or low-priority alarm signal specified in YY 0709.
The ventilator that is powered by DC power supply should activate a medium-priority alarm in
accordance with YY 0709 when the battery is low; and should activate a high-priority alarm in
accordance with YY 0709 before all power is lost.
If the ventilator is equipped with a DC power supply, a method for the operator to determine
the status of the DC power supply shall be provided.
5.3 Prevent mis-operation of the power switch
A method to prevent mis-operation of the power switch shall be provided.
5.4 Power cord
If the ventilator does not have a backup power supply, its equipped grid power cord shall be a
non-detachable power cord; or measures shall be provided to prevent the grid power cord from
accidentally falling off the ventilator.
6 Mechanical Strength
The ventilator that can be used for out-of-hospital transport shall be able to withstand the stress
caused by rough operation during working; and the ventilator shall continue to work during and
after the vibration impact test according to 21.6a), 21.6b) and 21.6aa) of YY 0600.3-2007.
7 Waterproof
The waterproof rating of ventilators that can be used for out-of-hospital transport shall be IPX4.
8 Fireproof
In normal or single fault conditions, combustible materials shall not be in both of the following
conditions at the same time:
--- the material temperature reaches the minimum ignition temperature;
--- the presence of oxidizing agents.
9 Cleaning, Disinfection or Sterilization
Accessories and components in the ventilator breathing system that may be contaminated with
rebreathing gases shall be removable for cleaning, disinfection, or sterilization.
If the label indicates that the product is sterile, the product should be sterilized by appropriate
and effective methods in accordance with the relevant descriptions in GB 18278.1, GB 18279.1,
GB 18280.1 and GB 18281.
Packaging systems without passing through the sterilized equipment shall maintain product
cleanliness prior to sterilization, use; and minimize the risk of bacterial contamination.
(0.1kPa + 10% of the setting value).
11.6 Sigh
If the ventilator has a sigh function, the tidal volume during sighing shall be no less than 1.5
times the tidal volume during non-sighing.
11.7 Measurement of airway pressure
A device to measure the airway pressure at the patient's connection port shall be provided; and
the accuracy of the pressure measuring device shall be ± (2% full scale + 8% actual reading).
NOTE: The actual measurement can be carried out at any part o...
Share











